2015 ext escalade service manuals. Ohaus Adventurer AX4201 4200g x 0.1g with Internal Calibration The new model from OHAUS® strikes the perfect balance between inventive features. ADVENTURER SL BALANCES. The information contained in this manual is believed to be accurate at the time of publication, but Ohaus. Corporation assumes.

5 PARTS INFORMATION If you require replacement parts or would like to purchase accessories, inside the U.S., please call Ohaus Corporation toll-free at (800) 526-0659 and from outside the U.S., (973) 377-9000. An Ohaus Product Parts Specialist will be available to help you.

REPLACEMENT PARTS Part No. AC Adapters: US/Japan 100 V 11103743 US 100-120V 11103741 Europe 220-240V 11103740 UK 220-240V 11103742 Table version 220-240V Reqs.

Ohaus adventurer scale manual

Line cord, Aus. 11103745 Line cord for above 76199-01 Table version 220-240V Reqs.

Ohaus Adventurer Pro Manual

Line cord, US 11103745 Line cord for above 6569-00 Euro/Germany, Belgium, France 220-240V 11103744 Pan for 65g, 110g, 210g 3.5' (9cm) dia. 12106739 Pan for 150g, 310g, 3.9' (10cm) dia. 12105372 Pan for 510 to 4100g, 7.1' (18cm) 12105370 ACCESSORIES Part No. 20g Calibration Mass 49024-11 50g Calibration Mass 49054-11 100g Calibration Mass 49015-11 200g Calibration Mass 49025-11 500g Calibration Mass 49055-11 1kg Calibration Mass 49016-11 2kg Calibration Mass 49026-11 Security Device 76288-01 Impact printer SF42 In-Use Cover 400302-010 Error Codes When a problem occurs using the balance, the display will indi- cate an error code. Review the listed codes and follow instruc- tions to correct the problem. Err 2.0 Stability error.

Check if balance is located near vibrat- ing equipment or if air currents are affecting it. Err 3.0 Calibration error (wrong mass). Incorrect or no cali- bration mass used when performing calibration proce- dure.

Scale

Error will flash momentarily, then balance will use previous calibration data. Recalibrate correctly. Err 4 Invalid data checksum in type or adjust data. Return balance for servicing. Err 6.0 Parts counting error - Average piece weight is less than 1d.

Balance shows error then exits parts counting. Err 8.4 Over or under load. Sample being weighed exceeds the capacity of the balance.

If error occurs when the sample is within the balance capacity, balance may be incorrectly calibrated. An underload such as the pan off of the balance could also display Err 8.4.

Recali- brate the balance. Err 9 Internal data error. Return balance for servicing.

Err 9.8 Invalid checksum in calibration or setup data. The balance may need recalibration, particularly linearity calibration. If the error persists after recalibration, the balance must be serviced. Capacity 65g (1) 110g (1) 210g (1) 150g (1) 210g (1) 310g (1) 410g (1) 510g 1500g 2100g 3100g 4100g Readability 0.1mg 0.001g 0.01g 0.1g Weighing units/modes g, kg, mg, ct, N, lb, oz, ozt, gn, dwt, momme, mesghal, 3 taels, ticals, Parts Counting Repeatability (Std. Dev.) 0.1mg 0.001g 0.01g 0.1g Linearity ±0.2mg ±0.3mg ±0.002g ±0.01g ±0.02g ±0.1g Tare Range Full capacity by subtraction Stabilization Time (seconds) 3 Seconds Sensitivity Drift 4ppm/°C 10ppm/°C Operating Temp Range 50° to 86° F/10° to 30° C Power Requirements 8 - 14.5V 50/60Hz 6VA or 9.5 - 20V 6W Calibration External Digital Display (in/cm) LCD (1.2/3 high) Pan Size (in/cm) 3.5/9 dia.

Draft shield clearance height 9.3/23.6 6.8/17.25 above pan (in/cm) Dimensions (WxHxD) (in/cm) 8.5x14.3x13.5/21.7x36.3x34.3 8.5x12.3x13.5/21.7x31x34.3 8.5x4.3x13.5/21.7x11x34.3 Net Wt (lb/kg) 12.5/5.7 10/4.5 8.5/3.9 Shipping Wt (lb/kg) 21/9.5 16/7.3 14/6.4 Item No. AR0640 AR1140 AR2140 AR1530 AR2130 AR3130 AR4130 AR5120 ARA520 ARB120 ARC120 ARD110 SPECIFICATIONS Notes 1. 65g to 310g analytical balances include a three door draft shield. Some models are not available in all countries.